Instruction Sheet
012-10287B
Ethanol Sensor
PS-2194
Included Parts
1. Amplifier
2. Probe
3. PTFE tape for membrane replacement (not pictured)
Additional Part Required
PASPORT interface or datalogger
Related Parts
Replacement probe (PASCO part 003-10338)
Additional PTFE tape (720-062)
Quick Start
1. Connect the amplifier to your PASPORT interface.
2. Connect the probe to the amplifier.
3. If you are using a computer, connect the PASPORT interface
to it and start DataStudio.
4. Ensure that the membrane is installed and intact. There
should be no tears or discoloration.
5. Place the end of the probe in the gas above a solution con-
taining less than 3% ethanol.
6. Press or click the start button to begin recording data.
The probe stabilizes after about 10 minutes of warm up time.
Introduction
The Ethanol Sensor measures the concentration of ethanol in a
gas or indirectly in a liquid. The sensors probe contains a mate-
rial whose resistance decreases in the presence of ethanol. The
sensor translates this resistance into an ethanol concentration
measurement, which is recorded and displayed by the connected
computer or datalogger. The Ethanol Sensor is well-suited for
continuous recording and discrete measurements.
Precautions
Do not expose the probe to gas containing more than 3%
ethanol, including the gas above alcoholic beverages.
Higher ethanol concentrations will deplete the sensing ele-
ment. Note that the gas above a liquid contains about the
same concentration of ethanol as the liquid itself.
Ensure that the membrane is installed and intact before
using the probe. The membrane prevents the sensing element
from getting wet. If the sensing element accidentally gets
wet, rinse and dry it quickly. Note that the probe may be
degraded or destroyed.
Do not immerse the probe. The membrane (if it is in good
condition) will protect the probe in case of accidental immer-
sion. However, the probe is designed for use in gas only.
Connecting the Parts
1. Plug the amplifiers PASPORT connector into any port of a
PASPORT interface or datalogger.
2. Connect the probe to the amplifier.
Connecting the probe and interface
3. If you are using a computer, connect the PASPORT interface
to it and start DataStudio.
1
2
to PASPORT
interface
probe
Vue de la page 0
1 2 3

Résumé du contenu

Page 1 - Ethanol Sensor

®Instruction Sheet012-10287BEthanol SensorPS-2194Included Parts1. Amplifier2. Probe3. PTFE tape for membrane replacement (not pictured)Additional Part

Page 2

®Ethanol Sensor PS-21942Warm-upThe probe contains a heating element. After connecting the sen-sor to a powered PASPORT interface, wait about 10 minute

Page 3 - Technical Support

®Ethanol Sensor PS-21943Sensor Response versus TemperatureThe graph below shows how changing temperature affects the sensor. The probe was placed in

Commentaires sur ces manuels

Pas de commentaire