Pasco PS-2147 High Resolution pH-ORP-ISE Amplifier Manuel d'utilisateur

Naviguer en ligne ou télécharger Manuel d'utilisateur pour Équipement Pasco PS-2147 High Resolution pH-ORP-ISE Amplifier. PASCO PS-2147 High Resolution pH-ORP-ISE Amplifier User Manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 4
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 0
®
PS-2147
The PS-2147 Sensor is a multi-purpose millivolt (mV) meter that accepts standard
BNC connectors and works with both PASCO and industry standard ion selective
electrodes (ISE), including pH and ORP. The sensor is capable of measuring
temperature using PASCO’s thermistor-based temperature probes.
pH/ORP/ISE/Temp Quick Start
PASPORT
TM
interface (USB Link, PowerLink, Xplorer, etc.) with USB-compatible
computer or a PS-2000 Xplorer
EZscreen or DataStudio
®
software (version 1.8.5 or later)
1. Connect the PASPORT interface to a USB port on your computer or to a USB hub.
2. Connect an electrode to the BNC port and a temperature probe to the 3.5 mm plug
input on the sensor box.
3. Connect the PS-2147 sensor to the PASPORT interface. (If using an Xplorer in the
classroom, connect the Xplorer cable to your computer.)
4. The software launches when it detects a PASPORT sensor. Select a point of entry.
Sensor Specifications:
pH:
Range: 0 to 14
Accuracy:
±
0.1
Resolution: 0.001
ORP and ISE:
Range: -2000 mV to +2000 mV
Accuracy:
±
2.0 mV
Resolution: 0.05 mV
Temperature:
Range: -35
°
C to +135
°
C
Accuracy:
±
0.5
°
C
Resolution: 0.0025
°
C
Probe Specifications:
Fast Response Temperature
(PS-2135):
Range: -10
°
C to +70
°
C
Accuracy:
±
0.5
°
C
Resolution: 0.0025
°
C
Construction: A 10K thermistor is
housed in a water proof PVC tip. The
thermistor is connected to a 3.5 mm
stereo connector with a 4-foot PVC
insulated wire.
pH electrode (699-085):
Range: 0 to 12
Accuracy: See pH electrode calibration
on card 2A.
Resolution: 0.001
Temperature range: 0
°
C to +80
°
C
Construction: Double junction, electrode
with three-foot cable and BNC connector.
Reference is gel-filled.
800-772-8700 • 916-786-3800 [email protected] • www.pasco.com
012-08636B
1
2
3
High Resolution pH/ORP/ISE Amplifier
with Temperature Sensor
CARD 1A
4
Vue de la page 0
1 2 3 4

Résumé du contenu

Page 1 - Probe Specifications:

®PS-2147The PS-2147 Sensor is a multi-purpose millivolt (mV) meter that accepts standard BNC connectors and works with both PASCO and industry standar

Page 2 - Measure the temp. change:

DataStudio/DS Lite Tasks: Procedure:Open the Setup window:On the main toolbar, click the Setup button.Change measurements: In the Setup window, go t

Page 3 - A Practical Approach

Calibrating the pH Electrode: A Practical Approach CARD 2ACalibrating the ISE/ORP ElectrodesTypically, pH electrodes are sensitive to the condit

Page 4 - Sensor Usage Tips

CAUTION: Do not place the PS-2147 pH/ORP/ISE Sensor case in liquids or expose to chemicals. The connectors and internal electronic circuit may be da