Model No. TD-8580A Sample Data
7
Sample Data
DataStudio files containing these data can be found on the included CD-ROM.
Infrared Light Intensity versus Temperature
The graphs above show I versus T (above left) and I
versus T
4
(above right) for the black surface. The
curve fit applied to I versus T is ; note
that the curve intersects the horizontal axis at
, the approximate temperature of
the sensor.
The graphs of I versus T for the different surfaces of
the cube (right) show that the black and white surfaces
have similar emissivities in the IR region of the spec-
trum, and the emissivity of the polished surface is
much lower.
Reflected Visible Light and Emitted Infrared Light
These graphs show visible light intensity versus posi-
tion (top right) and IR light intensity versus position
(bottom right) at 92 °C. In the visible region of the
spectrum, the cavity is darker than the surrounding
cube surface; in the IR region the cavity is brighter.
Resistance to Temperature Conversion
To find the temperature (T) measured in kelvin based on the resistance (R) of the
embedded thermistor measured in k, use the formula: